SphygmoCor CPV

AtCor Medical, Australia

SphygmoCor_CPV
  • The SphygmoCor CPV is a time-consuming, user-dependent applanation tonometric method with ECG gating for PWVao measurement (27).

  • Its use requires trained personnel.

  • Existing cardiovascular outcome studies (33,34).

  • The device uses validated generalized transfer function (GTF) (28).

  • Invasively validated AIXao and SBPao (28,29,30).

  • PWVao is also validated invasively, however the limit of agreement according to Bland-Altman plot turned to be high -3.8 m/s (-2SD) and 4.1 m/s (+2SD) (31).

Green Yes.png

Carotid-femoral Pulse Wave Velocity (31,32)

Green Yes.png

Central Aortic Systolic Blood Pressure (28,29,30)

Green Yes.png

Aortic Augmentationindex (28)

Relevant publications of SphygmoCor CPV